15
en
URT/AR
• Operation
General
The URT/AR measures temperatures using Pt100 or Ni120 RTD
sensors and produces an industry standard, analogue signal
proportional to the temperature. It also has two alarm channels and a
display that shows the measured temperature.
Reviewing the setup
For review mode, press the Å and ì keys simultaneously.
Press the Å and ì keys to move through the setup sequence (see
table). Greyed items in the setup table are not shown in review mode.
A 10 s timeout applies for review mode or you can press to quit.
Note: You can use review mode while the instrument is in service and
the analogue and alarm outputs will still operate correctly.
RTD Temperature
sensor
Analogue Output
Alarm 1
Alarm 2
Power Supply
°C